Efficient Measurement of Superconducting Resonators

ORAL

Abstract

S-parameter measurements of high-Q superconducting resonators at single-photon drive powers often require significant averaging with associated long acquisition time. We have developed a procedure for optimizing the frequency sweep-plan of the measurement, and found that an appropriate choice of frequencies has a significant impact on its efficiency. An optimized sweep-plan design offers up to a factor of two reduction in the variance of extracted parameters, in comparison to a linear sweep-plan having the same total acquisition time. We experimentally compare the performance of the optimized and linear sweeps in measurements of high-Q aluminum CPW resonators.

Authors

  • Steven Sendelbach

    • Northrop Grumman
  • Micah Stoutimore

    • Northrop Grumman
  • Josh Strong

    • Northrop Grumman
  • Ofer Naaman

    • Northrop Grumman
  • Brooks Campbell

    • UC Santa Barbara
  • John Martinis

    • UC Santa Barbara