ORISS: A compact isomer and isobar separator for study of exotic decays

ORAL

Abstract

A compact isobar and isomer spectrometer and separator ORISS (Oak Ridge Isomer Separator and Spectrometer), based on the multi-pass-time-of-flight principle, is being constructed. A mass resolving power of 110,000 (fwhm) and a transmission of 50{\%} have been achieved as a spectrometer with an off-line ion source with large emittance. As a separator, molecules of N$_{2}$ and CO with a mass difference of 1/2500 or 10.433 MeV were separated with ToF peaks corresponding to a mass resolution of 40,000. For injection of radioactive ions into ORISS and to further improve its mass resolution, we have constructed cooler/buncher RF quadrupoles and demonstrated a bunch width of 9 ns (fwhm) and a transmission of 75 -- 80 {\%}. With this bunch width, ORISS can achieve a mass resolution of $\sim $ 400,000 and will be able to separate nuclei or isomers with a mass difference of 1/200,000, corresponding to 470 keV at mass A=100. At present, the quadrupoles are being integrated into the ORISS system. ORISS will be used for decay spectroscopy to provide isotopically pure samples of exotic species around $^{100}$Sn and of neutron rich nuclei. In addition, ORISS will allow a fast and efficient search for isomers within an entire isobaric chain.

Authors

  • A. Piechaczek

    • UNIRIB, ORAU, Oak Ridge, TN 37830
  • V. Shchepunov

    • UNIRIB, ORAU, Oak Ridge, TN 37830
  • H.K. Carter

    • UNIRIB, ORAU, Oak Ridge, TN 37830
  • J.C. Batchelder

    • UNIRIB, ORAU, Oak Ridge, TN 37830
  • E.F. Zganjar

    • Louisiana State University, Baton Rouge, LA 70803