A CsI detector array for NDTGamma test measurements

ORAL

Abstract

The goal of the NDTGamma experiment is to measure the directional asymmetry in gamma-ray emission from the reaction n + D $\to$ T + $\gamma$ (6.2 MeV). A 4x4 array of recycled CsI scintillators and photomultipliers tube (PMT) has been constructed for initial tests at the LANSCE FP12 beamline. Each detector was tested with different PMTs with a Co-60 gamma-ray source to calibrate the detectors and optimize high voltage and digital spectroscopy parameters. The signals were analyzed using both CAEN and low-cost Red Pitaya data acquisition systems. I will present the calibrations results and initial data from the FP12 tests.

*This material is based upon work supported by the U.S. Department of Energy, Office of Science, Office of Nuclear Physics under Award Number DE-SC0014622